Diabetes is a chronic disease that affects 347 million people world-wide. The disease is characterised by
insufficient or absent insulin production and secretion and/or insulin resistance, and the consequences are acute
and late-diabetic complications. Evidence from the Diabetes Control and Complications Trial suggests that
intensive insulin therapy delays the onset and slows the progression of late-diabetic complications. This beneficial
effect, however, comes at the expense of an increase in the number of acute hypoglycaemic events, which
hampers the therapeutic compliance because people with diabetes are afraid of hypoglycaemia. Measures to
detect hypoglycaemia, thereby enabling prevention, would be a possible solution to maintain intensive insulin
therapy without increasing the number of hypoglycaemic events. Self-monitoring of blood glucose typically results
in 3-4 blood glucose measurements pr. day, which is not enough to detect all hypoglycaemic events. On the other
hand, a measurement every 5 minutes with continuous glucose monitoring provides a sufficient amount of
information to detect hypoglycaemia. Unfortunately, this technology suffers from inaccuracy, especially in the
hypoglycaemic range, due to physiological delay and a delay caused by filter routines. Researchers have for
decades worked on the problem of inaccuracy of continuous glucose monitoring, and the devices have improved
significantly. Nevertheless, the measuring devices have unacceptable inaccuracies resulting in an unacceptable
number of false alerts.
The research described in this thesis utilises pattern classification approaches to optimise the hypoglycaemia
detection of continuous glucose monitoring. A large number of features from the continuous glucose monitoring
signal and insulin injection were systematically extracted, and then the dimension was reduced with SEPCOR
and Forward Selection. Using Support Vector Machines, each continuous glucose monitoring reading was
classified as hypoglycaemic or non-hypoglycaemic based on concurrent blood glucose readings. This approach
was used to develop a retrospective algorithm and a real-time algorithm using both historic and future data and
only historic data, respectively. Both algorithms managed to detect 100% of the hypoglycaemic events of the
dataset, with only one false positive. By comparison, the continuous glucose monitoring device alone detected
only 2/3 of the events, but with zero false positives. These results, while promising, should be generalised
through training and testing of the algorithms on several datasets, including datasets with spontaneous
hypoglycaemic events.